LAB AUTOCLAVES

PORTABLE AUTOCLAVES

Constructions:

Stainless Steel :
-Seamless die pressed body, lid, flange, fly screw, cross pin, heater cover X stand - all stainless steel (304 grade).
-The flange is argon welded to the body.

Aluminium :
-Die cast body and lid. External powder coated.

Fittings:
-Lid is fitted with pressure gauge, safety valve (set at 15 psi), extra safety fusible plug (set at approx. 21 psi) and manual exhaust valve.

Electric:
-Energy efficient tubular heater.
-Standard accessories:
-Spare fusible plug and spanner.

AUTOCLAVES-MINI-AUTOMATIC

Constructions:

Stainless Steel :
-Stainless steel body, lid, flange, fly screw, cross pin, heater cover X stand - all stainless steel (304 grade).
-The flange is argon welded to the body.

Fittings:
-Lid is fitted with pressure gauge, safety valve (set at 17 psi) and manual exhaust valve.

Control:
-The unit incorporates an electronic blind temperature control (btc) preset at 1210 C (15psi).
-BTC is internally connected to an electronic blind timer (with end cycle buzzer and auto shut off) preset at 20 minutes.Control
Once, purging is performed (manually), the cycle is automatic till end.
-As the control of pressure is by switching ON and OFF the heater and not venting the excess steam, energy is saved and the -Autoclave can be kept in an air conditioned room also.

Electric:
-Special custom made heater.
-Standard accessories
-Stainless steel dressing drum of 18 Ø x 18 cm deep (7” Ø x 7”).

Construction:
-Internal chamber, false bottom, trays and reservoir made of stainless steel.
-Door of cast aluminium duly spray finished with external M S channel reinforcement for single lever locking.
-External body fabricated from M S.

Temperatures:
-Sterilizing temperature (pressure) user selectable between 121oC (15 psi) and 134oC (30 psi) for model # 7710 and fixed at 121oC (15 psi) for # 7720.
-Drying temperature is factory preset.

Purging (Free steaming):
-Automatic purging by a controlled solenoid valve for saturated steam.

Timers:
-Two sequentially operating, digital timers for sterilizing & drying.
-End cycle buzzer with auto shut off.
-Electrical & Heater
-Incorporates a fuse for safety.
-Incoloy sheathed heater.
-Pressure cycle starts only after door is properly closed, but by pass switch provided for faster or independent drying with door slightly open.

Process:
-Once water is filled to the required level, the material loaded, the two timers set, power switched ON and door tightened, the process is fully automatic to the end.
-Automatic purging (free steaming) taken care of by a solenoid valve.
-The selected temperature (pressure) 121oC or 134oC (if applicable) is attained. AFTER attaining the set temperature the countdown of the sterilizing timer starts. During this period, the controller controls the temperature (pressure).
-After the sterilizing time is over, water and steam are automatically condensed back into the reservoir.
-The drying timer starts its cycle.
-A buzzer sounds and shuts off the whole system at the end.